Severity
Moderate
FDA Devices recall · Reported March 9, 2022
Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magn…
Beckman Coulter Inc. recalled AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850 - a moderate-severity action.
AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850 was recalled by Beckman Coulter Inc. in March 9, 2022. Reason: Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in t…. Check the official notice for the remedy. Verify recall #Z-0692-2022 with the FDA Devices before acting.
The recall
Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0692-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0692-2022) was formally reported on March 9, 2022, with the manufacturer initiating the action on December 8, 2021. It is classified under Moderate severity (Class II), with a current status of Terminated. Beckman Coulter Inc. is listed as the recalling firm, operating out of Brea, CA. Federal records list the affected scope as 181,776 Kits.
The documented reason for this recall is: Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is onl… Distribution data in the federal record shows the product reached: US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
181,776 Kits
Related Recalls
6
6 from same agency
AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0692-2022 |
| Date reported | March 9, 2022 |
| Date initiated | December 8, 2021 |
| Recalling firm | Beckman Coulter Inc. |
| Firm location | Brea, CA |
| Affected scope | 181,776 Kits |
| Distribution | US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massach… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 9, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.