PlainRecalls
FDA Devices Moderate Class II Ongoing

30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10; Spiros, Closed Male Luer w/Red Cap, 5 Units, REF: CH2000SC-5; Spinning Spiros Closed Male Luer, Pu

Reported: December 30, 2020 Initiated: August 31, 2020 #Z-0702-2021

Product Description

30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10; Spiros, Closed Male Luer w/Red Cap, 5 Units, REF: CH2000SC-5; Spinning Spiros Closed Male Luer, Purple Cap, REF: CH2000S-PC; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC; Spiros, Purple Cap, Drop-In Priming Cap, REF: CH2000S-PCP; Spinning Spiros Closed Male Luer w/Red Cap, REF: CH3515; Spinning Spiros Closed Male Luer, REF: Z7072

Reason for Recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Details

Recalling Firm
ICU Medical, Inc.
Units Affected
1856578
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Location
San Clemente, CA

Frequently Asked Questions

What product was recalled?
30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10; Spiros, Closed Male Luer w/Red Cap, 5 Units, REF: CH2000SC-5; Spinning Spiros Closed Male Luer, Purple Cap, REF: CH2000S-PC; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC; Spiros, Purple Cap, Drop-In Priming Cap, REF: CH2000S-PCP; Spinning Spiros Closed Male Luer w/Red Cap, REF: CH3515; Spinning Spiros Closed Male Luer, REF: Z7072. Recalled by ICU Medical, Inc.. Units affected: 1856578.
Why was this product recalled?
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 30, 2020. Severity: Moderate. Recall number: Z-0702-2021.