OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.
Reported: December 25, 2024 Initiated: November 21, 2024 #Z-0703-2025
Product Description
OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.
Reason for Recall
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Details
- Recalling Firm
- GE OEC Medical Systems, Inc
- Units Affected
- 84 systems
- Distribution
- U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
- Location
- Salt Lake City, UT
Frequently Asked Questions
What product was recalled? ▼
OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.. Recalled by GE OEC Medical Systems, Inc. Units affected: 84 systems.
Why was this product recalled? ▼
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 25, 2024. Severity: Moderate. Recall number: Z-0703-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…