PlainRecalls
FDA Devices Moderate Class II Ongoing

Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set

Reported: December 30, 2020 Initiated: August 31, 2020 #Z-0706-2021

Product Description

Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch¿tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC

Reason for Recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Details

Recalling Firm
ICU Medical, Inc.
Units Affected
217522
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Location
San Clemente, CA

Frequently Asked Questions

What product was recalled?
Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch¿tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC. Recalled by ICU Medical, Inc.. Units affected: 217522.
Why was this product recalled?
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 30, 2020. Severity: Moderate. Recall number: Z-0706-2021.