FDA Devices recall · Reported January 17, 2024

XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-XXXXX; d) XS6 WAM TTL KIT IEC XML BRAZIL, XSCRIBE-5AX-XXBRZ; e) XS6 WAM TTL SYS Z200 AHA XML, XSCRIBE-6AA-AAAAA; f) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-ADBXX; g) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAAA; h) XS6 WAM TTL SYS Z200 PCRT AHA SVR, XSCRIBE-6AA-AFAAC; i) X

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Recall #
Z-0706-2024
Affected scope
2985 units
Initiated
December 8, 2023
Verify with FDA Devices →
View my saved recalls

Baxter Healthcare Corporation recalled XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) … - a moderate-severity action.

XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) … was recalled by Baxter Healthcare Corporation in January 17, 2024. Reason: Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) …. Check the official notice for the remedy. Verify recall #Z-0706-2024 with the FDA Devices before acting.

The recall

Baxter Healthcare Corporation issued this moderate-severity FDA Devices recall-Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) ….

Moderate
severity level
3K units
affected scope
Class II
classification
January 17, 2024
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0706-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-XXXXX; d) XS6 WAM TTL KIT IEC XML BRAZIL, XSCRIBE-5AX-XXBRZ; e) XS6 WAM TTL SYS Z200 AHA XML, XSCRIBE-6AA-AAAAA; f) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-ADBXX; g) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAAA; h) XS6 WAM TTL SYS Z200 PCRT AHA SVR, XSCRIBE-6AA-AFAAC; i) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAXA; j) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-AFAXB; k) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AA-BCAAB; l) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AA-BXAAA; m) XS6 WAM TTL SYS AHA DICOM, XSCRIBE-6AA-BXAAB; n) XS6 WAM TTL SYS AHA, XSCRIBE-6AA-BXAXX; o) XS6 WAM TTL SYS Z200 BCRT IEC XML, XSCRIBE-6AA-CDBAA; p) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-CDBAB; q) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AA-CDBAC; r) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-CDBAX; s) XS6 WAM TTL SYS BCRT AHA DICOM, XSCRIBE-6AA-CEAAB; t) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AA-CEBAB; u) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-DDBAB; v) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-DGAAA; w) XS6 WAM TTL SYS Z200 PCRT IEC DICOM, XSCRIBE-6AA-DGBAB; x) XS6 WAM TTL SYS Z200 PCRT IEC, XSCRIBE-6AA-DGBAX; y) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAAA; z) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-XFAAB; aa) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAXA; bb) XS6 WAM TTL SYS Z200 PCRT IEC XML, XSCRIBE-6AA-XFBXA; cc) XS6 WAM TTL SYS BCRT AHA, XSCRIBE-6AB-BEAXX; dd) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-BXAXA; ee) XS6 WAM TTL SYS Z200 BCRT AHA XML, XSCRIBE-6AB-CDAAA; ff) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AB-CDAAB; gg) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAA; hh) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAB; ii) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AB-CDBAC; jj) XS6 WAM TTL SYS BCRT AHA XML, XSCRIBE-6AB-CEAAA; kk) XS6 WAM TTL SYS BCRT IEC, XSCRIBE-6AB-CEBAX; ll) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-DDBAB; mm) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AB-DDBAX; nn) XS6 WAM TTL SYS BCRT IEC XML, XSCRIBE-6AB-DEBAA; oo) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AB-DEBAB; pp) XS6 WAM TTL SYS BCRT IEC SVR, XSCRIBE-6AB-DEBAC; qq) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAAA; rr) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAXA; ss) XS6 WAM TTLSYS Z200 BCRT AHA XML, XSCRIBE-6AB-XCAAA; tt) XS6 WAM TTLKIT Z200 AHA DICOM, XSCRIBE-6AC-AAAAB; uu) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAAX; vv) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAXX; ww) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAAX; xx) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAXX; yy) XS6 WAM TTL KIT BCRT AHA XML, XSCRIBE-6AC-AEAAA; zz) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-AEAXX; aaa) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAAA; bbb) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAXA; ccc) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-AXAXX; ddd) XS6 WAM TTL KIT Z200 BCRT AHA DICOM, XSCRIBE-6AC-BDAAB; eee) XS6 WAM TTL KIT BCRT AHA SVR, XSCRIBE-6AC-BEAAC; fff) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-BEAXX; ggg) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-BXAXA; hhh) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-BXAXX; iii) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-CDBAA; jjj) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-CDBAB; kkk) XS6 WAM TTL KIT BCRT IEC, XSCRIBE-6AC-CEBAX; lll) XS6 WAM TTL KIT IEC XML, XSCRIBE-6AC-CXBAA; mmm) XS6 WAM TTL KIT IEC DICOM, XSCRIBE-6AC-CXBAB; nnn) XS6 WAM TTL KIT Z200 IEC XML, XSCRIBE-6AC-DBBXA; ooo) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DCBXA; ppp) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DDBAA; qqq) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBAB; rrr) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBXB; sss) XS6 WAM TTL KIT BCRT IEC XML, XSCRIBE-6AC-DEBAA; ttt) XS6 WAM TTL KIT BCRT IEC

Reason for recall

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0706-2024
Date reported January 17, 2024
Date initiated December 8, 2023
Recalling firm Baxter Healthcare Corporation
Firm location Deerfield, IL
Affected scope 2985 units
Distribution United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croati…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

2985 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0706-2024) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-XXXXX; d) XS6 WAM TTL KIT IEC XML BRAZIL, XSCRIBE-5AX-XXBRZ; e) XS6 WAM TTL SYS Z200 AHA XML, XSCRIBE-6AA-AAAAA; f) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-ADBXX; g) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAAA; h) XS6 WAM TTL SYS Z200 PCRT AHA SVR, XSCRIBE-6AA-AFAAC; i) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAXA; j) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-AFAXB; k) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AA-BCAAB; l) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AA-BXAAA; m) XS6 WAM TTL SYS AHA DICOM, XSCRIBE-6AA-BXAAB; n) XS6 WAM TTL SYS AHA, XSCRIBE-6AA-BXAXX; o) XS6 WAM TTL SYS Z200 BCRT IEC XML, XSCRIBE-6AA-CDBAA; p) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-CDBAB; q) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AA-CDBAC; r) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-CDBAX; s) XS6 WAM TTL SYS BCRT AHA DICOM, XSCRIBE-6AA-CEAAB; t) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AA-CEBAB; u) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-DDBAB; v) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-DGAAA; w) XS6 WAM TTL SYS Z200 PCRT IEC DICOM, XSCRIBE-6AA-DGBAB; x) XS6 WAM TTL SYS Z200 PCRT IEC, XSCRIBE-6AA-DGBAX; y) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAAA; z) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-XFAAB; aa) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAXA; bb) XS6 WAM TTL SYS Z200 PCRT IEC XML, XSCRIBE-6AA-XFBXA; cc) XS6 WAM TTL SYS BCRT AHA, XSCRIBE-6AB-BEAXX; dd) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-BXAXA; ee) XS6 WAM TTL SYS Z200 BCRT AHA XML, XSCRIBE-6AB-CDAAA; ff) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AB-CDAAB; gg) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAA; hh) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAB; ii) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AB-CDBAC; jj) XS6 WAM TTL SYS BCRT AHA XML, XSCRIBE-6AB-CEAAA; kk) XS6 WAM TTL SYS BCRT IEC, XSCRIBE-6AB-CEBAX; ll) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-DDBAB; mm) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AB-DDBAX; nn) XS6 WAM TTL SYS BCRT IEC XML, XSCRIBE-6AB-DEBAA; oo) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AB-DEBAB; pp) XS6 WAM TTL SYS BCRT IEC SVR, XSCRIBE-6AB-DEBAC; qq) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAAA; rr) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAXA; ss) XS6 WAM TTLSYS Z200 BCRT AHA XML, XSCRIBE-6AB-XCAAA; tt) XS6 WAM TTLKIT Z200 AHA DICOM, XSCRIBE-6AC-AAAAB; uu) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAAX; vv) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAXX; ww) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAAX; xx) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAXX; yy) XS6 WAM TTL KIT BCRT AHA XML, XSCRIBE-6AC-AEAAA; zz) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-AEAXX; aaa) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAAA; bbb) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAXA; ccc) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-AXAXX; ddd) XS6 WAM TTL KIT Z200 BCRT AHA DICOM, XSCRIBE-6AC-BDAAB; eee) XS6 WAM TTL KIT BCRT AHA SVR, XSCRIBE-6AC-BEAAC; fff) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-BEAXX; ggg) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-BXAXA; hhh) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-BXAXX; iii) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-CDBAA; jjj) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-CDBAB; kkk) XS6 WAM TTL KIT BCRT IEC, XSCRIBE-6AC-CEBAX; lll) XS6 WAM TTL KIT IEC XML, XSCRIBE-6AC-CXBAA; mmm) XS6 WAM TTL KIT IEC DICOM, XSCRIBE-6AC-CXBAB; nnn) XS6 WAM TTL KIT Z200 IEC XML, XSCRIBE-6AC-DBBXA; ooo) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DCBXA; ppp) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DDBAA; qqq) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBAB; rrr) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBXB; sss) XS6 WAM TTL KIT BCRT IEC XML, XSCRIBE-6AC-DEBAA; ttt) XS6 WAM TTL KIT BCRT IEC. Recalled by Baxter Healthcare Corporation. Units affected: 2985 units.
Why was this product recalled?
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 17, 2024. Severity: Moderate. Recall number: Z-0706-2024.
Where was the recalled product distributed?
Distribution: United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Argentina, Bahamas, Bolivia, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, and Trinidad and Tobago..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported January 17, 2024. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026