PlainRecalls
FDA Devices Moderate Class II Ongoing

QMS Tacrolimus Calibrators, Product code 10015573

Reported: January 17, 2024 Initiated: November 21, 2023 #Z-0712-2024

Product Description

QMS Tacrolimus Calibrators, Product code 10015573

Reason for Recall

Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.

Details

Units Affected
10 sales units US; 1 sales unit OUS
Distribution
Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to Chile.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
QMS Tacrolimus Calibrators, Product code 10015573. Recalled by Ortho-Clinical Diagnostics, Inc.. Units affected: 10 sales units US; 1 sales unit OUS.
Why was this product recalled?
Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 17, 2024. Severity: Moderate. Recall number: Z-0712-2024.