FDA Devices recall · Reported January 24, 2024
Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may re…
Philips North America recalled Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cab… - a moderate-severity action.
Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cab… was recalled by Philips North America in January 24, 2024. Reason: If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the …. Check the official notice for the remedy. Verify recall #Z-0719-2024 with the FDA Devices before acting.
The recall
Philips North America issued this moderate-severity FDA Devices recall-If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the ….
- Moderate
- severity level
- Class II
- classification
- January 24, 2024
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0719-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871
Reason for recall
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0719-2024 |
| Date reported | January 24, 2024 |
| Date initiated | December 7, 2023 |
| Recalling firm | Philips North America |
| Firm location | Cambridge, MA |
| Affected scope | 55 US; 1029 ROW |
| Distribution | Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0719-2024) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.