Severity
Moderate
FDA Devices recall · Reported March 16, 2022
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R …
Philips North America Llc recalled Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chille… - a moderate-severity action.
Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chille… was recalled by Philips North America Llc in March 16, 2022. Reason: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. I…. Check the official notice for the remedy. Verify recall #Z-0725-2022 with the FDA Devices before acting.
The recall
Philips North America Llc issued this moderate-severity FDA Devices recall-Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. I….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0725-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0725-2022) was formally reported on March 16, 2022, with the manufacturer initiating the action on November 24, 2021. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips North America Llc is listed as the recalling firm, operating out of Cambridge, MA. Federal records list the affected scope as 1 unit.
The documented reason for this recall is: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, w… Distribution data in the federal record shows the product reached: US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Ne…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1 unit
Related Recalls
6
3 from same agency
Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0725-2022 |
| Date reported | March 16, 2022 |
| Date initiated | November 24, 2021 |
| Recalling firm | Philips North America Llc |
| Firm location | Cambridge, MA |
| Affected scope | 1 unit |
| Distribution | US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippin… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 16, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.