FDA Devices recall · Reported January 1, 2020
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent fo
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI…
Cardinal Health Inc. recalled S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The … - a moderate-severity action.
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The … was recalled by Cardinal Health Inc. in January 1, 2020. Reason: Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information…. Check the official notice for the remedy. Verify recall #Z-0728-2020 with the FDA Devices before acting.
The recall
Cardinal Health Inc. issued this moderate-severity FDA Devices recall-Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information….
- Moderate
- severity level
- 267K units
- affected scope
- Class II
- classification
- January 1, 2020
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0728-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope (see the Recall Profile below) is large enough to require multi-state distribution tracking. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.
Reason for recall
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0728-2020 |
| Date reported | January 1, 2020 |
| Date initiated | November 14, 2019 |
| Recalling firm | Cardinal Health Inc. |
| Firm location | Dublin, OH |
| Affected scope | 267,247 units total (262,641 OUS, 4,606 US) |
| Distribution | Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0728-2020) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026