Severity
Moderate
FDA Devices recall · Reported December 28, 2022
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be i…
Ge Medical Systems, LLC recalled B125M Patient Monitor, REF 6160000-004-XXXXXX - a moderate-severity action.
B125M Patient Monitor, REF 6160000-004-XXXXXX was recalled by Ge Medical Systems, LLC in December 28, 2022. Reason: GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit…. Check the official notice for the remedy. Verify recall #Z-0732-2023 with the FDA Devices before acting.
The recall
Ge Medical Systems, LLC issued this moderate-severity FDA Devices recall-GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0732-2023 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0732-2023) was formally reported on December 28, 2022, with the manufacturer initiating the action on November 16, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Ge Medical Systems, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 5962.
The documented reason for this recall is: GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that i… Distribution data in the federal record shows the product reached: Worldwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
5962
Related Recalls
6
3 from same agency
B125M Patient Monitor, REF 6160000-004-XXXXXX
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0732-2023 |
| Date reported | December 28, 2022 |
| Date initiated | November 16, 2022 |
| Recalling firm | Ge Medical Systems, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 5962 |
| Distribution | Worldwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 28, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.