LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for
Product Description
LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model¿ SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.
Reason for Recall
The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
Details
- Recalling Firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Units Affected
- 11
- Distribution
- US Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
- Location
- Norderstedt
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11