epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Reported: January 1, 2025 Initiated: November 22, 2024 #Z-0736-2025 670 units units

Siemens Healthcare Diagnostics, Inc. issued this FDA Devices recall on January 1, 2025. Classified as Moderate severity (Class II). Approximately 670 units units are affected. The recall was issued because: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occ…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0736-2025) was formally reported on January 1, 2025, with the manufacturer initiating the action on November 22, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of East Walpole, MA. Federal records indicate 670 units units are affected.

The documented reason for this recall is: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a… Distribution data in the federal record shows the product reached: Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt E…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

670 units

Related Recalls

6

6 from same agency

Product Description

epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Reason for Recall

Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.

Details

Units Affected
670 units
Distribution
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
Location
East Walpole, MA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0736-2025
Date reported January 1, 2025
Date initiated November 22, 2024
Recalling firm Siemens Healthcare Diagnostics, Inc.
Units affected 670 units
Distribution Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland Franc…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

670 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 670 units.
Why was this product recalled?
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 1, 2025. Severity: Moderate. Recall number: Z-0736-2025.
Where was the recalled product distributed?
Distribution: Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0736-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).