FDA Devices Moderate Class II Terminated

STA SYSTEM CONROL N + P (ref. 00678)

Reported: January 1, 2020 Initiated: November 18, 2019 #Z-0738-2020

Product Description

STA SYSTEM CONROL N + P (ref. 00678)

Reason for Recall

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.