Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology.

Reported: March 23, 2022 Initiated: February 3, 2022 #Z-0749-2022

Product Description

Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology.

Reason for Recall

Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.

Details

Recalling Firm
Abbott Laboratories, Inc
Units Affected
306 units
Distribution
Worldwide distribution: US (nationwide) including states of: AL, CA, GA, IL, IN, MN, MO, NC, OK, PA, RI, SC, TN, TX, WA, WI, MD and countries (OUS) including of: Australia, Austria, Bahrain, Belgium, Brazil, china, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Jordan Kuwait and Mexico.
Location
Irving, TX

Frequently Asked Questions

What product was recalled?
Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology.. Recalled by Abbott Laboratories, Inc. Units affected: 306 units.
Why was this product recalled?
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 23, 2022. Severity: Moderate. Recall number: Z-0749-2022.