FDA Devices recall · Reported February 17, 2016
When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the so…
Toshiba American Medical Systems Inc recalled INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems - a moderate-severity action.
INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems was recalled by Toshiba American Medical Systems Inc in February 17, 2016. Reason: When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the …. Check the official notice for the remedy. Verify recall #Z-0752-2016 with the FDA Devices before acting.
The recall
Toshiba American Medical Systems Inc issued this moderate-severity FDA Devices recall-When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0752-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0752-2016) was formally reported on February 17, 2016, with the manufacturer initiating the action on August 14, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Toshiba American Medical Systems Inc is listed as the recalling firm, operating out of Tustin, CA. Federal records list the affected scope as 4.
The documented reason for this recall is: When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defin… Distribution data in the federal record shows the product reached: US Nationwide Distribution to OH and NY. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4
Related Recalls
6
3 from same agency
INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems
When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0752-2016 |
| Date reported | February 17, 2016 |
| Date initiated | August 14, 2015 |
| Recalling firm | Toshiba American Medical Systems Inc |
| Firm location | Tustin, CA |
| Affected scope | 4 |
| Distribution | US Nationwide Distribution to OH and NY |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 17, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.