Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDR

Reported: January 24, 2024 Initiated: November 8, 2023 #Z-0758-2024 236793 units in total units

Getinge Usa Sales Inc issued this FDA Devices recall on January 24, 2024. Classified as Moderate severity (Class II). Approximately 236793 units in total units are affected. The recall was issued because: A potential for a light system to fall in the operating room.. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0758-2024) was formally reported on January 24, 2024, with the manufacturer initiating the action on November 8, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Getinge Usa Sales Inc is listed as the recalling firm, operating out of Wayne, NJ. Federal records indicate 236793 units in total units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: A potential for a light system to fall in the operating room. Distribution data in the federal record shows the product reached: US Nationwide. Global Distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 2 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

236793 units in total

Related Recalls

6

6 from same agency

Product Description

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS

Reason for Recall

A potential for a light system to fall in the operating room.

Details

Recalling Firm
Getinge Usa Sales Inc
Units Affected
236793 units in total
Distribution
US Nationwide. Global Distribution.
Location
Wayne, NJ

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0758-2024
Date reported January 24, 2024
Date initiated November 8, 2023
Recalling firm Getinge Usa Sales Inc
Units affected 236793 units in total
Distribution US Nationwide. Global Distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

236793 units in total units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS. Recalled by Getinge Usa Sales Inc. Units affected: 236793 units in total.
Why was this product recalled?
A potential for a light system to fall in the operating room.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 24, 2024. Severity: Moderate. Recall number: Z-0758-2024.
Where was the recalled product distributed?
Distribution: US Nationwide. Global Distribution..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0758-2024) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).