PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported January 24, 2024

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDR

A potential for a light system to fall in the operating room.

Recall #
Z-0758-2024
Affected scope
236793 units in total
Initiated
November 8, 2023
Verify with FDA Devices →
View my saved recalls

Getinge Usa Sales Inc recalled Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model… - a moderate-severity action.

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model… was recalled by Getinge Usa Sales Inc in January 24, 2024. Reason: A potential for a light system to fall in the operating room.. Check the official notice for the remedy. Verify recall #Z-0758-2024 with the FDA Devices before acting.

The recall

Getinge Usa Sales Inc issued this moderate-severity FDA Devices recall-A potential for a light system to fall in the operating room..

Moderate
severity level
237K units
affected scope
Class II
classification
January 24, 2024
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0758-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope (see the Recall Profile below) is large enough to require multi-state distribution tracking. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS

Reason for recall

A potential for a light system to fall in the operating room.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0758-2024
Date reported January 24, 2024
Date initiated November 8, 2023
Recalling firm Getinge Usa Sales Inc
Firm location Wayne, NJ
Affected scope 236793 units in total
Distribution US Nationwide. Global Distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

236793 units in total units affected - multi-state distribution scale.

Regional (<10K units) -
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0758-2024) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS. Recalled by Getinge Usa Sales Inc. Units affected: 236793 units in total.
Why was this product recalled?
A potential for a light system to fall in the operating room.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 24, 2024. Severity: Moderate. Recall number: Z-0758-2024.
Where was the recalled product distributed?
Distribution: US Nationwide. Global Distribution..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported January 24, 2024. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026