i-STAT cTnI cartridge

Reported: January 8, 2020 Initiated: October 16, 2019 #Z-0761-2020

Product Description

i-STAT cTnI cartridge

Reason for Recall

i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.

Details

Recalling Firm
Abbott Point Of Care Inc.
Units Affected
3,465,875
Distribution
US: AK AL AZ CA CT DC FL GA HI IL IN KS KY MD MI MS NC NE NH NM NV NY OH OK OR PA SC SD TN TX VA WA WV WY
Location
Princeton, NJ

Frequently Asked Questions

What product was recalled?
i-STAT cTnI cartridge. Recalled by Abbott Point Of Care Inc.. Units affected: 3,465,875.
Why was this product recalled?
i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 8, 2020. Severity: Moderate. Recall number: Z-0761-2020.