Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.

Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adju…

Recall #
Z-0765-2016
Affected scope
2 subject to correction and removal
Initiated
November 3, 2015
Verify with FDA Devices →
View my saved recalls

Philips Electronics North America Corporation recalled Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications. — a moderate-severity action.

Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications. was recalled by Philips Electronics North America Corporation in February 17, 2016. Reason: Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagneti…. Check the official notice for the remedy. Verify recall #Z-0765-2016 with the FDA Devices before acting.

The recall

Philips Electronics North America Corporation issued this moderate-severity FDA Devices recall — Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagneti….

Moderate
severity level
Class II
classification
February 17, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0765-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0765-2016) was formally reported on February 17, 2016, with the manufacturer initiating the action on November 3, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Electronics North America Corporation is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 2 subject to correction and removal.

The documented reason for this recall is: Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is… Distribution data in the federal record shows the product reached: Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

2 subject to correction and removal

Related Recalls

6

6 from same agency

Product description

Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.

Reason for recall

Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0765-2016
Date reported February 17, 2016
Date initiated November 3, 2015
Recalling firm Philips Electronics North America Corporation
Firm location Andover, MA
Affected scope 2 subject to correction and removal
Distribution Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0765-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.. Recalled by Philips Electronics North America Corporation. Units affected: 2 subject to correction and removal.
Why was this product recalled?
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Which agency issued this recall?
This recall was issued by the FDA Devices on February 17, 2016. Severity: Moderate. Recall number: Z-0765-2016.
Where was the recalled product distributed?
Distribution: Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0765-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported February 17, 2016.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.