FDA Devices Low Class III Terminated

BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamli

Reported: February 13, 2019 Initiated: January 2, 2019 #Z-0792-2019

Product Description

BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.

Reason for Recall

A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.

Details

Units Affected
3 devices
Distribution
US Distribution in MD.
Location
San Jose, CA

Frequently Asked Questions

What product was recalled?
BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.. Recalled by Becton, Dickinson and Company, BD Biosciences. Units affected: 3 devices.
Why was this product recalled?
A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 13, 2019. Severity: Low. Recall number: Z-0792-2019.