MEDTRONIC CARELINK 2090 programmer

Reported: February 13, 2019 Initiated: January 17, 2019 #Z-0795-2019

Product Description

MEDTRONIC CARELINK 2090 programmer

Reason for Recall

There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).

Details

Units Affected
33007 units
Distribution
US, Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
Location
Mounds View, MN

Frequently Asked Questions

What product was recalled?
MEDTRONIC CARELINK 2090 programmer. Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). Units affected: 33007 units.
Why was this product recalled?
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
Which agency issued this recall?
This recall was issued by the FDA Devices on February 13, 2019. Severity: Moderate. Recall number: Z-0795-2019.