FDA Devices recall · Reported January 29, 2014

GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to

Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respira…

Recall #
Z-0808-2014
Affected scope
1,948 (1,834 US; 114 OUS)
Initiated
December 5, 2013
Verify with FDA Devices →
View my saved recalls

Ge Healthcare, LLC recalled GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Modul… - a moderate-severity action.

GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Modul… was recalled by Ge Healthcare, LLC in January 29, 2014. Reason: Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on…. Check the official notice for the remedy. Verify recall #Z-0808-2014 with the FDA Devices before acting.

The recall

Ge Healthcare, LLC issued this moderate-severity FDA Devices recall-Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on….

Moderate
severity level
Class II
classification
January 29, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0808-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. The TRAM module functions as a transport monitor during transport, and in the bedside monitor during ICU or OR monitoring. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. TRAM modules acquire, process and store information regarding these parameters. TRAM modules incorporate different monitoring capabilities based on their configuration. K011000: The TRAM 2001 Module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The TRAM 2001 Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric, and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM 2001 Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM 2001 Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport. K053121: The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.

Reason for recall

Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0808-2014
Date reported January 29, 2014
Date initiated December 5, 2013
Recalling firm Ge Healthcare, LLC
Firm location Waukesha, WI
Affected scope 1,948 (1,834 US; 114 OUS)
Distribution Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0808-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. The TRAM module functions as a transport monitor during transport, and in the bedside monitor during ICU or OR monitoring. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. TRAM modules acquire, process and store information regarding these parameters. TRAM modules incorporate different monitoring capabilities based on their configuration. K011000: The TRAM 2001 Module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The TRAM 2001 Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric, and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM 2001 Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM 2001 Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport. K053121: The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.. Recalled by Ge Healthcare, LLC. Units affected: 1,948 (1,834 US; 114 OUS).
Why was this product recalled?
Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 29, 2014. Severity: Moderate. Recall number: Z-0808-2014.
Where was the recalled product distributed?
Distribution: Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported January 29, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026