BIOMET TRAUMA Drill Bit, Tap, Guide Pin: [DVR Drill Bit, Item # 212000022, DVR 2.2MM DRILL BIT], [ALPS Drill Bit, Item # 214227160, 2.7X160MM CALIBRATED DRILL], [ALPS Drill Bit, Item # 824273100, 4.5X150MM DRILL], [ALPS Drill Bit, Item # 829029170, 2.5X140MM DRILL], [ALPS Drill Bit, Item # 829032170, 3.5X140MM DRILL], [ALPS Tap, Item # 824150170, 3.5X140MM CORTICAL TAP], [ALPS Guide Pin, Item # 829516150, 1.6X150MM BAYONET TIP GUIDE PIN], [ALPS Guide Pin, Item # 829516151, 1.6X150MM THREADED TIP
Product Description
BIOMET TRAUMA Drill Bit, Tap, Guide Pin: [DVR Drill Bit, Item # 212000022, DVR 2.2MM DRILL BIT], [ALPS Drill Bit, Item # 214227160, 2.7X160MM CALIBRATED DRILL], [ALPS Drill Bit, Item # 824273100, 4.5X150MM DRILL], [ALPS Drill Bit, Item # 829029170, 2.5X140MM DRILL], [ALPS Drill Bit, Item # 829032170, 3.5X140MM DRILL], [ALPS Tap, Item # 824150170, 3.5X140MM CORTICAL TAP], [ALPS Guide Pin, Item # 829516150, 1.6X150MM BAYONET TIP GUIDE PIN], [ALPS Guide Pin, Item # 829516151, 1.6X150MM THREADED TIP GUIDE PIN]. Instruments are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.
Reason for Recall
Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.
Details
- Recalling Firm
- Biomet, Inc.
- Units Affected
- 2594
- Distribution
- Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands
- Location
- Warsaw, IN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11