GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Product Description
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Reason for Recall
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Details
- Recalling Firm
- GE Medical Systems, LLC
- Units Affected
- 120 units
- Distribution
- Worldwide
- Location
- Waukesha, WI
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11