Severity
Moderate
FDA Devices recall · Reported December 24, 2014
When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left V…
Philips Ultrasound, Inc. recalled Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ… - a moderate-severity action.
Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ… was recalled by Philips Ultrasound, Inc. in December 24, 2014. Reason: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications,…. Check the official notice for the remedy. Verify recall #Z-0815-2015 with the FDA Devices before acting.
The recall
Philips Ultrasound, Inc. issued this moderate-severity FDA Devices recall-When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications,….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0815-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0815-2015) was formally reported on December 24, 2014, with the manufacturer initiating the action on November 5, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Ultrasound, Inc. is listed as the recalling firm, operating out of Bothell, WA. Federal records list the affected scope as additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128 outside US).
The documented reason for this recall is: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be… Distribution data in the federal record shows the product reached: Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countrie…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128 outside US)
Related Recalls
6
3 from same agency
Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ and/or aCMQ plug-ins installed with Philips Xcelera software. *********************************************************************************** Devices identified in the Field Safety Notice ( FSN MA-FCO79500316 ) QLAB Quantification Software. QLAB version 10.0, 10.1 and 10.1.1 with a2DQ and/or aCMQ plug-ins are affected. Catalog # 795041 and Part # 989605315491 Designed to view and quantify image data acquired on Philips Healthcare ultrasound products.
When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0815-2015 |
| Date reported | December 24, 2014 |
| Date initiated | November 5, 2014 |
| Recalling firm | Philips Ultrasound, Inc. |
| Firm location | Bothell, WA |
| Affected scope | additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128 outside US) |
| Distribution | Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emi… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 24, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.