Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

Reported: January 29, 2020 Initiated: April 19, 2019 #Z-0821-2020

Product Description

Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

Reason for Recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Details

Recalling Firm
Isopure Corp
Units Affected
15
Distribution
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
Location
Louisville, KY

Frequently Asked Questions

What product was recalled?
Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3. Recalled by Isopure Corp. Units affected: 15.
Why was this product recalled?
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.