Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Reported: January 29, 2020 Initiated: April 19, 2019 #Z-0823-2020

Product Description

Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Reason for Recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Details

Recalling Firm
Isopure Corp
Units Affected
8
Distribution
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
Location
Louisville, KY

Frequently Asked Questions

What product was recalled?
Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part). Recalled by Isopure Corp. Units affected: 8.
Why was this product recalled?
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.