FDA Devices recall · Reported December 24, 2014
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of…
Zimmer Manufacturing B.v. recalled Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, … - a moderate-severity action.
Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, … was recalled by Zimmer Manufacturing B.v. in December 24, 2014. Reason: Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified…. Check the official notice for the remedy. Verify recall #Z-0824-2015 with the FDA Devices before acting.
The recall
Zimmer Manufacturing B.v. issued this moderate-severity FDA Devices recall-Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0824-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0824-2015) was formally reported on December 24, 2014, with the manufacturer initiating the action on November 21, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Zimmer Manufacturing B.v. is listed as the recalling firm, operating out of N/A. Federal records list the affected scope as 6,377.
The documented reason for this recall is: Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. . Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and Internationally to Canada Australia Brazil Chile China Colombia Japan Malaysia Singapore South Korea Taiwan Albania Austria Belgium Bulgaria Canary Islands Czech Republic De…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
6,377
Related Recalls
6
3 from same agency
Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0824-2015 |
| Date reported | December 24, 2014 |
| Date initiated | November 21, 2014 |
| Recalling firm | Zimmer Manufacturing B.v. |
| Firm location | N/A |
| Affected scope | 6,377 |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Canada Australia Brazil Chile China Colombia Japan Malaysia Singapore South Korea Taiwan Albania Austria Belgium Bulgaria Canary Islands Czech Republic Denmark Egypt Finland … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 24, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.