Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008
Reported: March 30, 2022 Initiated: February 14, 2022 #Z-0824-2022
Product Description
Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008
Reason for Recall
Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label.
Details
- Recalling Firm
- Cytocell Ltd.
- Units Affected
- 5 kits
- Distribution
- US Nationwide distribution in the states of CT, TX.
- Location
- Cambridge, N/A
Frequently Asked Questions
What product was recalled? ▼
Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008. Recalled by Cytocell Ltd.. Units affected: 5 kits.
Why was this product recalled? ▼
Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on March 30, 2022. Severity: Low. Recall number: Z-0824-2022.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11