Severity
Critical
FDA Devices recall · Reported February 27, 2013
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleev…
DePuy Orthopaedics, Inc. recalled LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LP… - a critical-severity action.
LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LP… was recalled by DePuy Orthopaedics, Inc. in February 27, 2013. Reason: DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the …. Check the official notice for the remedy. Verify recall #Z-0828-2013 with the FDA Devices before acting.
The recall
DePuy Orthopaedics, Inc. issued this critical-severity FDA Devices recall-DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0828-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0828-2013) was formally reported on February 27, 2013, with the manufacturer initiating the action on January 4, 2013. It is classified under Critical severity (Class I), with a current status of Terminated. DePuy Orthopaedics, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 162 total units (US) and 102 units total (OUS).
The documented reason for this recall is: DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connec… Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
162 total units (US) and 102 units total (OUS)
Related Recalls
6
3 from same agency
LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects.
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-0828-2013 |
| Date reported | February 27, 2013 |
| Date initiated | January 4, 2013 |
| Recalling firm | DePuy Orthopaedics, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 162 total units (US) and 102 units total (OUS) |
| Distribution | Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South Africa and Canada in… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 27, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.