PlainRecalls
FDA Devices Moderate Class II Terminated

Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.

Reported: February 5, 2014 Initiated: September 17, 2013 #Z-0834-2014

Product Description

Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.

Reason for Recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Details

Recalling Firm
DeRoyal Industries Inc
Units Affected
738 units
Distribution
Nationwide Distribution.
Location
New Tazewell, TN

Frequently Asked Questions

What product was recalled?
Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.. Recalled by DeRoyal Industries Inc. Units affected: 738 units.
Why was this product recalled?
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 5, 2014. Severity: Moderate. Recall number: Z-0834-2014.