Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Product Description
Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Reason for Recall
Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 1,949
- Distribution
- Worldwide Distribution -- Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, and United States.
- Location
- Newark, DE
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11