Severity
Moderate
FDA Devices recall · Reported April 6, 2022
Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the …
Siemens Medical Solutions USA, Inc recalled Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system - a moderate-severity action.
Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system was recalled by Siemens Medical Solutions USA, Inc in April 6, 2022. Reason: Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully aut…. Check the official notice for the remedy. Verify recall #Z-0841-2022 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc issued this moderate-severity FDA Devices recall-Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully aut….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0841-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0841-2022) was formally reported on April 6, 2022, with the manufacturer initiating the action on January 31, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Siemens Medical Solutions USA, Inc is listed as the recalling firm, operating out of Malvern, PA. Federal records list the affected scope as 70 Systems.
The documented reason for this recall is: Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If t… Distribution data in the federal record shows the product reached: US Nationwide.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
70 Systems
Related Recalls
6
6 from same agency
Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system
Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0841-2022 |
| Date reported | April 6, 2022 |
| Date initiated | January 31, 2022 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Firm location | Malvern, PA |
| Affected scope | 70 Systems |
| Distribution | US Nationwide. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 6, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.