Hanger-Bar 2P 45cm

Reported: April 6, 2022 Initiated: November 1, 2021 #Z-0851-2022

Product Description

Hanger-Bar 2P 45cm

Reason for Recall

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Details

Recalling Firm
Human Care USA, Inc.
Units Affected
10 devices
Distribution
U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.
Location
Austin, TX

Frequently Asked Questions

What product was recalled?
Hanger-Bar 2P 45cm. Recalled by Human Care USA, Inc.. Units affected: 10 devices.
Why was this product recalled?
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 6, 2022. Severity: Moderate. Recall number: Z-0851-2022.