Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.

Reported: February 27, 2013 Initiated: January 16, 2013 #Z-0866-2013

Product Description

Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.

Reason for Recall

A consumer complaint that a catheter was observed to be in several pieces under patient fluoroscopic visualization.

Details

Units Affected
387 units
Distribution
Worldwide Distribution - USA including the states of NJ, FL, GA, MO, AZ, UT and the country of Denmark
Location
South Jordan, UT

Frequently Asked Questions

What product was recalled?
Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.. Recalled by Merit Medical Systems, Inc.. Units affected: 387 units.
Why was this product recalled?
A consumer complaint that a catheter was observed to be in several pieces under patient fluoroscopic visualization.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 27, 2013. Severity: Moderate. Recall number: Z-0866-2013.