Philips, REF M5072A, INFANT/CHILD - SMART PADS CARTRIDGE (0-8 YEARS). Smart PADS CARTRIDGE , (0-8 YEARS) <55 lbs/25 kg, CE 0123, NON-STERILE, Rx only FOR USE WITH HEARSTART DEFIBRILLATORS

Reported: April 13, 2022 Initiated: February 4, 2022 #Z-0882-2022 331,712 UNITS units

Philips North America LLC issued this FDA Devices recall on April 13, 2022. Classified as Moderate severity (Class II). Approximately 331,712 UNITS units are affected. The recall was issued because: There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic l…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0882-2022) was formally reported on April 13, 2022, with the manufacturer initiating the action on February 4, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips North America LLC is listed as the recalling firm, operating out of Bothell, WA. Federal records indicate 331,712 UNITS units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may s… Distribution data in the federal record shows the product reached: Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Mala…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 4 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

331,712 UNITS

Related Recalls

6

6 from same agency

Product Description

Philips, REF M5072A, INFANT/CHILD - SMART PADS CARTRIDGE (0-8 YEARS). Smart PADS CARTRIDGE , (0-8 YEARS) <55 lbs/25 kg, CE 0123, NON-STERILE, Rx only FOR USE WITH HEARSTART DEFIBRILLATORS

Reason for Recall

There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.

Details

Recalling Firm
Philips North America LLC
Units Affected
331,712 UNITS
Distribution
Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, Chile, Ireland, Norway, Hong Kong, Argentina, Mexico, Bolivia, Croatia, French Guiana, Denmark, United Arab Emirates, Hungary, Belgium, Indonesia, Korea, Republic of, Bhutan, New Zealand, Bulgaria, Mauritius, Trinidad and Tobago, Israel, Guatemala, Mali, Algeria, Philippines, Malta, Curacao, Panama, Paraguay, Costa Rica, Colombia, Liechtenstein, Saudi Arabia, Monaco, Estonia, Senegal, Reunion, Luxembourg, Qatar, El Salvador, Namibia, Myanmar, Nigeria, Poland, Aruba, Greece, Martinique, Guadeloupe, Dominican Republic, China, Kosovo, Portugal, Bangladesh, Tanzania, Peru, Lithuania, New Caledonia, State of Palestine, Finland, Cambodia, Macao, Cyprus, Zambia, Viet Nam, Uruguay, Ecuador, Netherlands Antilles, Czech Republic, French Polynesia, Kenya, Cameroon, Guinea, Mayotte, and Brunei Darussalam.
Location
Bothell, WA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0882-2022
Date reported April 13, 2022
Date initiated February 4, 2022
Recalling firm Philips North America LLC
Units affected 331,712 UNITS
Distribution Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, C…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

331,712 UNITS units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Philips, REF M5072A, INFANT/CHILD - SMART PADS CARTRIDGE (0-8 YEARS). Smart PADS CARTRIDGE , (0-8 YEARS) <55 lbs/25 kg, CE 0123, NON-STERILE, Rx only FOR USE WITH HEARSTART DEFIBRILLATORS. Recalled by Philips North America LLC. Units affected: 331,712 UNITS.
Why was this product recalled?
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 13, 2022. Severity: Moderate. Recall number: Z-0882-2022.
Where was the recalled product distributed?
Distribution: Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, Chile, Ireland, Norway, Hong Kong, Argentina, Mexico, Bolivia, Croatia, French Guiana, Denmark, United Arab Emirates, Hungary, Belgium, Indonesia, Korea, Republic of, Bhutan, New Zealand, Bulgaria, Mauritius, Trinidad and Tobago, Israel, Guatemala, Mali, Algeria, Philippines, Malta, Curacao, Panama, Paraguay, Costa Rica, Colombia, Liechtenstein, Saudi Arabia, Monaco, Estonia, Senegal, Reunion, Luxembourg, Qatar, El Salvador, Namibia, Myanmar, Nigeria, Poland, Aruba, Greece, Martinique, Guadeloupe, Dominican Republic, China, Kosovo, Portugal, Bangladesh, Tanzania, Peru, Lithuania, New Caledonia, State of Palestine, Finland, Cambodia, Macao, Cyprus, Zambia, Viet Nam, Uruguay, Ecuador, Netherlands Antilles, Czech Republic, French Polynesia, Kenya, Cameroon, Guinea, Mayotte, and Brunei Darussalam..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0882-2022) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).