PlainRecalls
CriticalClass IOngoing

FDA Devices recall · Reported December 24, 2025

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event o…

Recall #
Z-0915-2026
Affected scope
203 units
Initiated
November 14, 2025
Verify with FDA Devices →
View my saved recalls

Ge Medical Systems China Co., Ltd. recalled GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceabl… — a critical-severity action.

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceabl… was recalled by Ge Medical Systems China Co., Ltd. in December 24, 2025. Reason: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power m…. Check the official notice for the remedy. Verify recall #Z-0915-2026 with the FDA Devices before acting.

The recall

Ge Medical Systems China Co., Ltd. issued this critical-severity FDA Devices recall — Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power m….

Critical
severity level
203 units
affected scope
Class I
classification
December 24, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0915-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0915-2026) was formally reported on December 24, 2025, with the manufacturer initiating the action on November 14, 2025. It is classified under Critical severity (Class I), with a current status of Ongoing. Ge Medical Systems China Co., Ltd. is listed as the recalling firm, operating out of Wuxi. Federal records list the affected scope as 203 units.

The documented reason for this recall is: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthes… Distribution data in the federal record shows the product reached: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Domini…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Critical

Affected scope

203 units

Related Recalls

6

6 from same agency

Product description

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

Reason for recall

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0915-2026
Date reported December 24, 2025
Date initiated November 14, 2025
Recalling firm Ge Medical Systems China Co., Ltd.
Firm location Wuxi
Affected scope 203 units
Distribution Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuado…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

203 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0915-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • This is a Class I (high-risk) recall — stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S. Recalled by Ge Medical Systems China Co., Ltd.. Units affected: 203 units.
Why was this product recalled?
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 24, 2025. Severity: Critical. Recall number: Z-0915-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0915-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported December 24, 2025.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.