SOMATOM Perspective, Computed tomography x-ray system
Reported: January 4, 2017 Initiated: November 22, 2016 #Z-0921-2017
Product Description
SOMATOM Perspective, Computed tomography x-ray system
Reason for Recall
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Details
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Units Affected
- 13 systems
- Distribution
- Distributed to: IL IL GA SC SD OK NY MI IN MD AR
- Location
- Malvern, PA
Frequently Asked Questions
What product was recalled? ▼
SOMATOM Perspective, Computed tomography x-ray system. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 13 systems.
Why was this product recalled? ▼
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on January 4, 2017. Severity: Moderate. Recall number: Z-0921-2017.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11