Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Reported: December 31, 2025 Initiated: November 18, 2025 #Z-0923-2026

Product Description

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Reason for Recall

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Details

Recalling Firm
AGFA Healthcare Corp.
Units Affected
35
Distribution
U.S. Nationwide distribution.
Location
Greenville, SC

Frequently Asked Questions

What product was recalled?
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800. Recalled by AGFA Healthcare Corp.. Units affected: 35.
Why was this product recalled?
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 31, 2025. Severity: Moderate. Recall number: Z-0923-2026.