Severity
Moderate
FDA Devices recall · Reported March 14, 2018
During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding me…
Medtronic Navigation, Inc.-Littleton recalled O-arm¿ O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000 The O… - a moderate-severity action.
O-arm¿ O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000 The O… was recalled by Medtronic Navigation, Inc.-Littleton in March 14, 2018. Reason: During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that …. Check the official notice for the remedy. Verify recall #Z-0927-2018 with the FDA Devices before acting.
The recall
Medtronic Navigation, Inc.-Littleton issued this moderate-severity FDA Devices recall-During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0927-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0927-2018) was formally reported on March 14, 2018, with the manufacturer initiating the action on September 30, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Medtronic Navigation, Inc.-Littleton is listed as the recalling firm, operating out of Littleton, MA. Federal records list the affected scope as 449.
The documented reason for this recall is: During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnectin… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, OH, OK, OR, P…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
449
Related Recalls
6
6 from same agency
O-arm¿ O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighting 60 lbs or greater and having an abdominal thickness of greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnecting process which may result in an electrical short and a shock to the field service technician.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0927-2018 |
| Date reported | March 14, 2018 |
| Date initiated | September 30, 2017 |
| Recalling firm | Medtronic Navigation, Inc.-Littleton |
| Firm location | Littleton, MA |
| Affected scope | 449 |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, V… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Two federal-filing-derived peer sets for this notice, both outside the same recalling firm and filing state so the neighborhoods are not firm or geographic containment.
Same archive severity grade (Moderate), closest by report date among other firms and states.
Same product category, closest by report date among other firms and states (deduped against the severity set).
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 14, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.