PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported January 14, 2015

Synthes Small Electric Drive (SED), part number: 05.001.175: The Small Electric Drive is designed for use in general traumatology, especially hand and food applications involving surgical procedures such as drilling, burring, reaming, pin and wire placement, cutting of bone and hard tissue.

The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start. Unintended start of the hand piece or wrong mode…

Recall #
Z-0931-2015
Affected scope
12
Initiated
November 6, 2014
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Synthes (USA) Products LLC recalled Synthes Small Electric Drive (SED), part number: 05.001.175: The Small Electric Drive is … - a moderate-severity action.

Synthes Small Electric Drive (SED), part number: 05.001.175: The Small Electric Drive is … was recalled by Synthes (USA) Products LLC in January 14, 2015. Reason: The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or un…. Check the official notice for the remedy. Verify recall #Z-0931-2015 with the FDA Devices before acting.

The recall

Synthes (USA) Products LLC issued this moderate-severity FDA Devices recall-The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or un….

Moderate
severity level
12 units
affected scope
Class II
classification
January 14, 2015
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0931-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0931-2015) was formally reported on January 14, 2015, with the manufacturer initiating the action on November 6, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Synthes (USA) Products LLC is listed as the recalling firm, operating out of West Chester, PA. Federal records list the affected scope as 12.

The documented reason for this recall is: The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start. Unintended start of the hand piece or wrong mode/direction could cause bone/soft tissue,… Distribution data in the federal record shows the product reached: US Nationwide Distribution to Florida only. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

12

Related Recalls

6

3 from same agency

Product description

Synthes Small Electric Drive (SED), part number: 05.001.175: The Small Electric Drive is designed for use in general traumatology, especially hand and food applications involving surgical procedures such as drilling, burring, reaming, pin and wire placement, cutting of bone and hard tissue.

Reason for recall

The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start. Unintended start of the hand piece or wrong mode/direction could cause bone/soft tissue, peripheral nerve, and /or vascular damage. Injuries may need surgical or medical intervention. Surgical delay and postoperative local infection are also potential risks.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0931-2015
Date reported January 14, 2015
Date initiated November 6, 2014
Recalling firm Synthes (USA) Products LLC
Firm location West Chester, PA
Affected scope 12
Distribution US Nationwide Distribution to Florida only

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

12 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0931-2015) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Synthes Small Electric Drive (SED), part number: 05.001.175: The Small Electric Drive is designed for use in general traumatology, especially hand and food applications involving surgical procedures such as drilling, burring, reaming, pin and wire placement, cutting of bone and hard tissue.. Recalled by Synthes (USA) Products LLC. Units affected: 12.
Why was this product recalled?
The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start. Unintended start of the hand piece or wrong mode/direction could cause bone/soft tissue, peripheral nerve, and /or vascular damage. Injuries may need surgical or medical intervention. Surgical delay and postoperative local infection are also potential risks.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 14, 2015. Severity: Moderate. Recall number: Z-0931-2015.
Where was the recalled product distributed?
Distribution: US Nationwide Distribution to Florida only.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0931-2015) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 14, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.