FDA Devices recall · Reported February 12, 2014
Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the pati…
Stryker Medical Division of Stryker Corporation recalled The Gaymar Medi-Therm device provides a means of regulating patient temperature by supply… - a moderate-severity action.
The Gaymar Medi-Therm device provides a means of regulating patient temperature by supply… was recalled by Stryker Medical Division of Stryker Corporation in February 12, 2014. Reason: Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature dis…. Check the official notice for the remedy. Verify recall #Z-0933-2014 with the FDA Devices before acting.
The recall
Stryker Medical Division of Stryker Corporation issued this moderate-severity FDA Devices recall-Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature dis….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0933-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0933-2014) was formally reported on February 12, 2014, with the manufacturer initiating the action on December 27, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Stryker Medical Division of Stryker Corporation is listed as the recalling firm, operating out of Portage, MI. Federal records list the affected scope as 1002 units.
The documented reason for this recall is: Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with … Distribution data in the federal record shows the product reached: nationwide and worldwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1002 units
Related Recalls
6
3 from same agency
The Gaymar Medi-Therm device provides a means of regulating patient temperature by supplying temperature-controlled water through a connector hose to accessory Gaymar Hyper/Hypothermia blanket(s)/body wrap(s). The blanket/body wrap provides an interface for heating or cooling the patient. Accessory YSI 400 series patient probe interfaces between the Medi-Therm and patient to sense patient temperature, which is displayed on the device's control panel. The Medi-Therm device controls output water temperature by mixing hot and cold water using hot and cold solenoid valves under microprocessor control. The device includes a circulating pump, heater and refrigeration system. The Medi-Therm controller, connector hose, blanket/body wrap and patient probe comprise the Medi-Therm Hyper/Hypothermia System.
Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with the set point. The water display will be blank during Auto mode. If the user wants to see the water temperature displayed and switches out of Auto mode to Manual mode, then back to
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0933-2014 |
| Date reported | February 12, 2014 |
| Date initiated | December 27, 2013 |
| Recalling firm | Stryker Medical Division of Stryker Corporation |
| Firm location | Portage, MI |
| Affected scope | 1002 units |
| Distribution | nationwide and worldwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 12, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.