FDA Devices recall · Reported March 6, 2019

The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.

Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed all…

Recall #
Z-0939-2019
Affected scope
1,252 Units
Initiated
December 20, 2018
Verify with FDA Devices →
View my saved recalls

Merge Healthcare, Inc. recalled The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digita… - a moderate-severity action.

The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digita… was recalled by Merge Healthcare, Inc. in March 6, 2019. Reason: Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowab…. Check the official notice for the remedy. Verify recall #Z-0939-2019 with the FDA Devices before acting.

The recall

Merge Healthcare, Inc. issued this moderate-severity FDA Devices recall-Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowab….

Moderate
severity level
1K units
affected scope
Class II
classification
March 6, 2019
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0939-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.

Reason for recall

Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0939-2019
Date reported March 6, 2019
Date initiated December 20, 2018
Recalling firm Merge Healthcare, Inc.
Firm location Hartland, WI
Affected scope 1,252 Units
Distribution Distribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1,252 Units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0939-2019) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.. Recalled by Merge Healthcare, Inc.. Units affected: 1,252 Units.
Why was this product recalled?
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
Which agency issued this recall?
This recall was issued by the FDA Devices on March 6, 2019. Severity: Moderate. Recall number: Z-0939-2019.
Where was the recalled product distributed?
Distribution: Distribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West , Virginia, Wisconsin. International distribution to Algeria, Austria, Belgium, Canada, Czech Republic, Denmark, Germany, Israel, Netherlands, Puerto Rico, United Kingdom..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported March 6, 2019. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026