INFINITY DUAL HEMO MCable Pod

Reported: March 21, 2018 Initiated: October 4, 2017 #Z-0942-2018

Product Description

INFINITY DUAL HEMO MCable Pod

Reason for Recall

Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.

Details

Units Affected
2,760
Distribution
Nationally
Location
Andover, MA