FDA Devices recall · Reported January 21, 2015
Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of t…
DePuy Orthopaedics, Inc. recalled The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replace… - a moderate-severity action.
The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replace… was recalled by DePuy Orthopaedics, Inc. in January 21, 2015. Reason: Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If…. Check the official notice for the remedy. Verify recall #Z-0953-2015 with the FDA Devices before acting.
The recall
DePuy Orthopaedics, Inc. issued this moderate-severity FDA Devices recall-Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0953-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0953-2015) was formally reported on January 21, 2015, with the manufacturer initiating the action on November 10, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. DePuy Orthopaedics, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 7,944.
The documented reason for this recall is: Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if relea… Distribution data in the federal record shows the product reached: Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
7,944
Related Recalls
6
3 from same agency
The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE instruments: keel punch, tibial tower, evaluation bullet, fixed bearing tibial impactor, rotating platform tibial impactor, and femoral impactor, for the purpose of impaction, extraction, and insertion applications at steps within the surgical procedure. Due to the frequency of its use, two impaction handles are supplied in each kit.
Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0953-2015 |
| Date reported | January 21, 2015 |
| Date initiated | November 10, 2014 |
| Recalling firm | DePuy Orthopaedics, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 7,944 |
| Distribution | Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabi… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 21, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.