CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device
Reported: February 8, 2023 Initiated: December 9, 2022 #Z-0965-2023
Product Description
CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device
Reason for Recall
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Details
- Recalling Firm
- Smiths Medical ASD Inc.
- Units Affected
- 214668 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN
- Location
- Minneapolis, MN
Frequently Asked Questions
What product was recalled? ▼
CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device. Recalled by Smiths Medical ASD Inc.. Units affected: 214668 units.
Why was this product recalled? ▼
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on February 8, 2023. Severity: Critical. Recall number: Z-0965-2023.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11