Artis Q zeego, Model Number 10848283

Reported: April 27, 2022 Initiated: February 23, 2022 #Z-0977-2022

Product Description

Artis Q zeego, Model Number 10848283

Reason for Recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Details

Units Affected
38 systems
Distribution
US Nationwide.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Artis Q zeego, Model Number 10848283. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 38 systems.
Why was this product recalled?
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".