Severity
Moderate
FDA Devices recall · Reported January 18, 2017
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning Syst…
Varian Medical Systems, Inc. recalled Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowl… - a moderate-severity action.
Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowl… was recalled by Varian Medical Systems, Inc. in January 18, 2017. Reason: Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in con…. Check the official notice for the remedy. Verify recall #Z-0990-2017 with the FDA Devices before acting.
The recall
Varian Medical Systems, Inc. issued this moderate-severity FDA Devices recall-Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in con….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0990-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0990-2017) was formally reported on January 18, 2017, with the manufacturer initiating the action on December 2, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Varian Medical Systems, Inc. is listed as the recalling firm, operating out of Palo Alto, CA. Federal records list the affected scope as 42 devices are affected.
The documented reason for this recall is: Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain con… Distribution data in the federal record shows the product reached: US Distribution to the states of : NJ, TN. OR and FL.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
42 devices are affected
Related Recalls
6
3 from same agency
Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0990-2017 |
| Date reported | January 18, 2017 |
| Date initiated | December 2, 2016 |
| Recalling firm | Varian Medical Systems, Inc. |
| Firm location | Palo Alto, CA |
| Affected scope | 42 devices are affected |
| Distribution | US Distribution to the states of : NJ, TN. OR and FL. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 18, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.