FDA Devices recall · Reported December 31, 2025
Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG
Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experien…
Medtronic Neuromodulation recalled Restore Clinician Programmer Application (CP App), Model A71100 used with the following p… - a moderate-severity action.
Restore Clinician Programmer Application (CP App), Model A71100 used with the following p… was recalled by Medtronic Neuromodulation in December 31, 2025. Reason: Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, thi…. Check the official notice for the remedy. Verify recall #Z-0996-2026 with the FDA Devices before acting.
The recall
Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, thi….
- Moderate
- severity level
- 8K units
- affected scope
- Class II
- classification
- December 31, 2025
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0996-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG
Reason for recall
Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0996-2026 |
| Date reported | December 31, 2025 |
| Date initiated | November 20, 2025 |
| Recalling firm | Medtronic Neuromodulation |
| Firm location | Minneapolis, MN |
| Affected scope | 7667 devices |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denma… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0996-2026) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026