Ossur CB Resolve, Sterile, Part/ Description: 505300C/ CB RESOLVE STERILE UNIT SML; 505400C/ CB RESOLVE STERILE UNIT STD; 505500C/ CB Resolve Sterile Unit LRG; 515300C/ CB RESOLVE RING KIT SM-No App; 515400C/ CB RESOLVE RING KIT STD-No App; 515500C/ CB RESOLVE RING KIT LG-No App; 540C/ CB RESOLVE COMPONENT TRAY; 553CB/ CB RING TRAY SM; 554CB/ CB RING TRAY STD; 555CB/ CB RING TRAY LG; 505301C/ Kit Closed Back Halo Sm; 505401C/ Kit Closed Back Halo Md; 505501C/ Kit Closed Back Hal

Reported: March 20, 2019 Initiated: June 21, 2018 #Z-1000-2019

Product Description

Ossur CB Resolve, Sterile, Part/ Description: 505300C/ CB RESOLVE STERILE UNIT SML; 505400C/ CB RESOLVE STERILE UNIT STD; 505500C/ CB Resolve Sterile Unit LRG; 515300C/ CB RESOLVE RING KIT SM-No App; 515400C/ CB RESOLVE RING KIT STD-No App; 515500C/ CB RESOLVE RING KIT LG-No App; 540C/ CB RESOLVE COMPONENT TRAY; 553CB/ CB RING TRAY SM; 554CB/ CB RING TRAY STD; 555CB/ CB RING TRAY LG; 505301C/ Kit Closed Back Halo Sm; 505401C/ Kit Closed Back Halo Md; 505501C/ Kit Closed Back Halo Lg

Reason for Recall

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

Details

Recalling Firm
Ossur Americas
Units Affected
656
Distribution
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
Location
Foothill Ranch, CA

Frequently Asked Questions

What product was recalled?
Ossur CB Resolve, Sterile, Part/ Description: 505300C/ CB RESOLVE STERILE UNIT SML; 505400C/ CB RESOLVE STERILE UNIT STD; 505500C/ CB Resolve Sterile Unit LRG; 515300C/ CB RESOLVE RING KIT SM-No App; 515400C/ CB RESOLVE RING KIT STD-No App; 515500C/ CB RESOLVE RING KIT LG-No App; 540C/ CB RESOLVE COMPONENT TRAY; 553CB/ CB RING TRAY SM; 554CB/ CB RING TRAY STD; 555CB/ CB RING TRAY LG; 505301C/ Kit Closed Back Halo Sm; 505401C/ Kit Closed Back Halo Md; 505501C/ Kit Closed Back Halo Lg. Recalled by Ossur Americas. Units affected: 656.
Why was this product recalled?
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 20, 2019. Severity: Moderate. Recall number: Z-1000-2019.