Severity
Critical
FDA Devices recall · Reported January 28, 2026
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its sp…
Medline Industries, LP recalled MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO - a critical-severity action.
MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO was recalled by Medline Industries, LP in January 28, 2026. Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires.…. Check the official notice for the remedy. Verify recall #Z-1014-2026 with the FDA Devices before acting.
The recall
Medline Industries, LP issued this critical-severity FDA Devices recall-Medline Industries has identified a potential hazard involving the hand control pendant and associated wires.….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1014-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1014-2026) was formally reported on January 28, 2026, with the manufacturer initiating the action on November 26, 2025. It is classified under Critical severity (Class I), with a current status of Ongoing. Medline Industries, LP is listed as the recalling firm, operating out of Northfield, IL. Federal records list the affected scope as 22565 units.
The documented reason for this recall is: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an ob… Distribution data in the federal record shows the product reached: US, CA, PA, GUAM, VI. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices and 2 from Medline Industries, LP - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
22565 units
Related Recalls
6
3 from same agency
MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-1014-2026 |
| Date reported | January 28, 2026 |
| Date initiated | November 26, 2025 |
| Recalling firm | Medline Industries, LP |
| Firm location | Northfield, IL |
| Affected scope | 22565 units |
| Distribution | US, CA, PA, GUAM, VI |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 28, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.